About Microcyber
Microcyber Corporation is a high-tech enterprise founded by the Shenyang Institute of Automation, Chinese Academy of Sciences, engaged in the development, production, and application of networked control systems, industrial communication, and instrumentation. The company has undertaken national science and technology programs including the National Science and Technology Major Project and the 863 Program, and is the construction unit for the National Engineering Research Center for Networked Control Systems.
Microcyber developed the first internationally certified fieldbus protocol master stack and the first nationally certified fieldbus instrument, and co-developed China’s first industrial Ethernet protocol standard (EPA) and first industrial wireless communication standard (WIA-PA), both of which became IEC international standards. The company’s products and technologies have won two National Science and Technology Progress Awards and a National Science and Technology Invention Award, and have been exported to markets worldwide across more than 200 large-scale automation projects. Microcyber holds ISO 9001 and IATF 16949 (automotive) quality system certifications, and is a member of FieldComm Group and PROFIBUS & PROFINET International (PNO).
Visão Geral do Produto
The NCS-TT305 series Transmissor de temperatura HART is a high-precision, low-drift, high-reliability field device for process control. It integrates a broad set of functional modules, supporting both general temperature detection and more complex control strategies.
The transmitter uses digital signal processing, is compatible with a wide range of RTD and thermocouple sensors, and covers a wide measuring range. Its simple field-to-control-room interface helps reduce installation, commissioning, and maintenance costs.
The NCS-TT305 series supports 4-20mA + HART output with local LCD display, and is used across the petroleum, chemical, power, and metallurgy industries.
Especificações Chave
| Parâmetro | Especificação |
|---|---|
| Bus Interface | 4-20mA + HART 7 (supports Burst mode) |
| Versão do Protocolo | HART 7 |
| Bus Power Supply | Non-Ex: (13-45)VDC; Ex d: (13-45)VDC; Ex ia: (13-30)VDC |
| Wiring Method | 2/3/4-wire |
| Sinal de Entrada | Resistance: 0-500Ω, 0-2000Ω, 0-4000Ω, 0-40000Ω; RTD: Cu50, Cu50_GOST, Cu100, Cu100_GOST, Pt50_GOST, Callendar_Van_Dusen(Pt100), Pt100, Pt100_GOST, Pt200, Pt500, Pt1000, Ni50, Ni100, Ni120, Ni1000; mV: -100mV to 100mV; Thermocouple: B, E, J, K, N, R, S, T, L_GOST, A(W5ReW20Re), C(W5ReW26Re), D(W3ReW25Re), NiCrAuFe, CuAuFe |
| Número de Canais | Canal duplo |
| Precisão Atual | 0.05%; current temperature drift 0.005%/°C |
| Isolamento elétrico | 2kVAC/50VAC |
| Efeito da Tensão | ±0,0051 TP3T/V |
| Tempo de Resposta | 1.0s-1.65s, depending on sensor type and wiring method |
| Start-up Time | ≤10s |
| Ajuste de Amortecimento | Time constant 0-32s |
| CEM | ABNT NBR IEC 61326 |
| Ex Certification | IECEx: Ex ia IIC T6…T4 Ga; ATEX: II2G Ex db IIC T6…T4 Gb, II2D Ex tb IIIC T85°C…T110°C Db |
| Tela LCD | Suportado |
| Operação do Botão | Supported for local parameter setting |
| Temperatura de Operação | (-40 to 85)°C; (-55 to 85)°C optional |
| Faixa de umidade | (5-95)%RH |
| Temperatura de Armazenamento | -50 to 85°C |
| Vibration Resistance | 2-25Hz: ±1,6mm; 25-100Hz: ±4g |
| Classificação de proteção | IP67 (IP20 confirmed; IP67 certification in progress) |
| Dimensões | 139.7 × 111.9 × 112.0 mm |
| Material de Construção | Aluminum, polycarbonate |
| Peso | Approximately 1.67 kg |
Certifications and Approvals
Intrinsically Safe: Ex ia IIC T6…T4 Ga
- Ia: highest safety rating, suitable for Zone 0 (Ga)
- IIC: highest gas hazard group, covering gases such as hydrogen, acetylene, and carbon disulfide
- T6…T4: maximum surface temperature under 85°C (T6), 100°C (T5), or 135°C (T4)
- Reference standards: IEC 60079-0:2017, IEC 60079-11:2023
Explosion-proof (Flameproof): II2G Ex db IIC T6…T4 Gb; II2D Ex tb IIIC T85°C…T110°C Db
- Reference standards: EN IEC 60079-0:2018+A11:2024, EN 60079-1:2014+A11:2024
Segurança Funcional: The NCS-TT305 series is assessed for functional safety per IEC 61508. Full safety data, including SIL ratings by architecture, is available in the NCS-TT305H Series Intelligent Temperature Transmitter Functional Safety Manual.